Related Experiment Videos
High performance liquid chromatographic determination of fluvoxamine in human plasma
1Laboratoire de Toxicologie, Faculte de Pharmacie, Lille, France.
Biomedical Chromatography : BMC
|July 1, 1989
Abstract:
A method has been developed for the separation and measurement of fluvoxamine in human plasma by high performance liquid chromatography. The method uses metapramine as an internal standard and provides a limit of detection of about 1.5 ng/mL for fluvoxamine. At a concentration of 25 ng/mL, fluvoxamine could be measured within a coefficient of variation of +/- 5.82 of the mean and at 100 ng/mL within a CV of +/- 2.78 of the mean. The method has been applied to the analysis of plasma from patients undergoing fluvoxamine therapy.