Intracochlear inflammatory response to cochlear implant electrodes in humans
Mohammad Seyyedi1, Joseph B Nadol
1*Department of Otolaryngology, Massachusetts Eye and Ear Infirmary; and †Department of Otology and Laryngology, Harvard Medical School, Boston, Massachusetts, U.S.A.
Summary
Inflammation is common around cochlear implant electrodes, particularly at the cochleostomy site. This tissue response may be linked to insertion trauma, impacting implant performance.
Area of Science:
- Otolaryngology
- Biomaterials Science
- Pathology
Background:
- Cochlear implant electrodes are generally considered biocompatible.
- However, tissue reactions to electrodes can lead to complications like inflammation and implant failure.
Purpose of the Study:
- To evaluate the types and severity of tissue responses adjacent to multichannel cochlear implant electrodes.
- To investigate the relationship between electrode location and tissue reaction.
Main Methods:
- Histological analysis of 28 temporal bones with cochlear implants.
- Scoring of cellular responses (eosinophils, lymphocytes, giant cells, new bone, fibrosis) at three electrode segments.
- Statistical comparison using the Wilcoxon test.
Main Results:
- A granulomatous reaction was observed in 96.4% of temporal bones.
- Eosinophil infiltration, suggesting allergy, occurred in 25% of cases.
- Inflammatory response was significantly higher near the cochleostomy (p < 0.05).
Conclusions:
- Inflammatory response is a frequent finding post-cochlear implantation.
- The heightened response at the cochleostomy suggests insertion trauma as a contributing factor.