Accurate feedback of chest compression depth on a manikin on a soft surface with correction for total body
Stefanie G Beesems1, Rudolph W Koster1
1Academic Medical Center, Department of Cardiology, Amsterdam, The Netherlands.
Objective:
TrueCPR is a new real-time compression depth feedback device that measures changes in magnetic field strength between a back pad and a chest pad. We determined its accuracy with a manikin on a test bench and on various surfaces.
Methods:
First, calibration and accuracy of the manikin and TrueCPR was verified on a drill press. Then, manual chest compressions were given, on a firm surface and on a foam or air mattress, with feedback of the TrueCPR or Q-CPR accelerometer, to achieve a depth of 50mm. Compression depth measurements by the devices and the manikin were compared.
Results:
On a hard surface TrueCPR showed a systematic underestimation of 2-3mm in the drill press. Manual tests on a hard surface showed a slightly larger underestimation of 4.5mm. When guided by TrueCPR on a foam or air mattress, the TrueCPR measured a mean(±SD) chest compression depth of 52.0(±1.9)mm and 49.4(±2.6)mm respectively, while the manikin measured 54.4(±1.8)mm and 52.1(±1.4)mm, respectively (p<0.001). When guided by the Q-CPR accelerometer on a foam or air mattress, the accelerometer measured depth of 54.3(±3.6)mm and 56.0(±3.8)mm respectively, compared to the manikin 42.4(±2.3)mm and 34.9(±3.6)mm, respectively (p<0.001).
Conclusion:
TrueCPR measures depth precisely, independent of the stiffness of the surface upon which the CPR is being performed with a constant inaccuracy of <4.5mm. A sternum-only accelerometer substantially overestimates depth when performing CPR on a soft surface. Correction for body displacement on a soft surface is essential for accurate delivery of chest compressions within the recommended depth range.
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Body Water Content and Fluid Compartments
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of the Abdomen III: Palpation