Related Experiment Video
Updated: Apr 25, 2026

Noninvasive Determination of Vortex Formation Time Using Transesophageal Echocardiography During Cardiac Surgery
Published on: November 28, 2018
Echocardiography underestimates stroke volume and aortic valve area: implications for patients with small-area
Calvin W L Chin1, Hwan J Khaw2, Elton Luo2
1BHF/University Centre for Cardiovascular Science, University of Edinburgh, Edinburgh, United Kingdom; National Heart Center Singapore, Singapore.
Background:
Discordance between small aortic valve area (AVA; < 1.0 cm(2)) and low mean pressure gradient (MPG; < 40 mm Hg) affects a third of patients with moderate or severe aortic stenosis (AS). We hypothesized that this is largely due to inaccurate echocardiographic measurements of the left ventricular outflow tract area (LVOTarea) and stroke volume alongside inconsistencies in recommended thresholds.
Methods:
One hundred thirty-three patients with mild to severe AS and 33 control individuals underwent comprehensive echocardiography and cardiovascular magnetic resonance imaging (MRI). Stroke volume and LVOTarea were calculated using echocardiography and MRI, and the effects on AVA estimation were assessed. The relationship between AVA and MPG measurements was then modelled with nonlinear regression and consistent thresholds for these parameters calculated. Finally the effect of these modified AVA measurements and novel thresholds on the number of patients with small-area low-gradient AS was investigated.
Results:
Compared with MRI, echocardiography underestimated LVOTarea (n = 40; -0.7 cm(2); 95% confidence interval [CI], -2.6 to 1.3), stroke volumes (-6.5 mL/m(2); 95% CI, -28.9 to 16.0) and consequently, AVA (-0.23 cm(2); 95% CI, -1.01 to 0.59). Moreover, an AVA of 1.0 cm(2) corresponded to MPG of 24 mm Hg based on echocardiographic measurements and 37 mm Hg after correction with MRI-derived stroke volumes. Based on conventional measures, 56 patients had discordant small-area low-gradient AS. Using MRI-derived stroke volumes and the revised thresholds, a 48% reduction in discordance was observed (n = 29).
Conclusions:
Echocardiography underestimated LVOTarea, stroke volume, and therefore AVA, compared with MRI. The thresholds based on current guidelines were also inconsistent. In combination, these factors explain > 40% of patients with discordant small-area low-gradient AS.
Related Concept Videos
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Aortic Regurgitation I: Introduction