Related Experiment Video
Updated: Apr 25, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Successful percutaneous retrieval of a dislodged left atrial appendage occlusion device with double transseptal
Ngai-Yin Chan1, Chi-Chung Choy, Chun-Leung Lau
1Department of Medicine and Geriatrics, Princess Margaret Hospital, Kowloon, Hong Kong.
Abstract:
A 64-year-old woman underwent left atrial appendage occlusion with an Amplatzer Cardiac Plug device. Displacement of the device was detected on day 1 with transesophageal echocardiographic checking. The device became dislodged and flitted in the left atrium after unsuccessful retrieval with a snare and 12 Fr steerable transseptal sheath. A double transseptal sheath technique was then attempted. The flitting device was stabilized by one 12 Fr steerable transseptal sheath and successfully retrieved with a biopsy bioptome through another similar transseptal sheath. The patient suffered from no long-term sequelae.