Cone beam computed tomography and low-dose multislice computed tomography in orthodontics and dentistry: a
1Department of Orthodontics and Orofacial Orthopedics, Zahnklinik 3 - Kieferorthopädie, Universitätsklinikum Erlangen, Glückstr. 11, 91054, Erlangen, Germany, elisabeth.hofmann@uk-erlangen.de.
Objectives:
The goal of the present work was to assess various computed tomography (CT) systems in term of image quality and organ doses, namely five cone-beam CT (CBCT) scanners operated at standard settings and three multislice CT (MSCT) scanners operated at greatly dose-reduced settings.
Materials And Methods:
Radiographic volume scans were taken on a complete human cadaveric head specimen and the image quality of each was rated by four experienced examiners according to specific skeletal structures and bone-soft tissue interfaces. Radiation doses were captured by a head-and-neck phantom (Rando; Alderson Research Laboratories). Standard protocols were used for the CBCT scans. For the MSCT scans, tube voltage and current were adjusted to minimize radiation without compromising image quality.
Results:
Interobserver agreement was close to perfect, with iota coefficients of 0.931 (95% CI 0.807-0.978) between groups 1 and 2 and 0.959 (95% CI 0.869-1.000) between groups 1 and 3. Ratings of image quality in terms of skeletal-structure representation were slightly better for the CBCT than the MSCT scanners, although these differences were not statistically significant. The two groups of scanners applied considerably different organ doses: the lowest dose (0.03 mSv) was measured on the bone surface with a CBCT unit (Picasso Trio® from Vatech) and the highest dose (8.30 mSv) in the vicinity of the eye lens with another CBCT unit (3D Accuitomo 170® from J. Morita).
Conclusion:
The various systems tested offer similar imaging quality but demonstrated distinct differences in organ dose levels. The decision on which approach to take is not between CBCT and MSCT but rather between specific models and parameter settings. If these are optimized, MSCT images providing useful clinical information can be obtained at much reduced levels of radiation. Depending on the model and setting used, MSCT radiation levels may even be lower than during CBCT scans.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
09:10Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...