Abstract:
Mg alloys are receiving considerable attention for biomedical stents due to their combination of good mechanical properties and high biodegradability. Cold rolling is necessary to process Mg alloy tubes before final drawing and fabrication of the magnesium stents. In this paper, cold-rolled tubes were subjected to a cross-sectional reduction rate (ε) of up to 19.7%, and were further processed at various ratios of wall-thickness to diameter reduction (Q) from 0 to 2.24 with a constant ε of 19.7%. The results show that the cold-rolled tubes exhibited a rise in ultimate tensile strength (UTS), yield strength (YS), and a reduction in elongation as ε increased from 5.5% to 19.7%. UTS, YS and elongation decreased when Q was increased from 0 to 2.24. Mechanical twinning was observed and analysed. Extension twins increased with increasing ε and were almost saturated at a ε of 16.5%. Extension twins play an important role in determining the evolution of mechanical behaviour in the case of increasing ε, whilst contraction/double twins and secondary extension twins have a large effect on mechanical behaviour in the case of varying Q. The results indicate that the proportions and types of twins play a major role in determining the mechanical behaviour of the AZ31 tubes.