Related Experiment Video
Updated: Apr 25, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
"EcoRadiology"--pulling the plug on wasted energy in the radiology department
Colin J McCarthy1, Jan F Gerstenmaier1, Ailbhe C O' Neill1
1Department of Radiology, St. Vincent's University Hospital, Elm Park, Dublin 4, Ireland.
Rationale And Objectives:
We sought to evaluate the power consumption of various devices around the radiology department, audit our use of recycling, and review efforts by vendors to reduce the environmental impact of their products.
Materials And Methods:
Using a readily available power monitor, we calculated the power consumption of different devices around our department. In particular, we calculated the financial and environmental cost of leaving equipment on overnight and/or at weekends. When it was not possible to measure energy usage directly, we obtained and reviewed relevant technical manuals. We contacted vendors directly to document how the environmental impact of new technology and decommissioning aging technology is being tackled.
Results:
We found that 29 of 43 desktop computers and 25 of 27 picture archiving and communications system (PACS) reporting stations were left on needlessly overnight and/or at weekends, resulting in estimated electrical running costs while not in use of approximately $7253 per year, and CO2 emissions equivalent to the annual emissions of over 10 passenger cars. We discovered that none of our PACS reporting stations supported energy-saving modes such as "sleep" or "hibernate." Despite encouraging staff to turn off computers when not in use, a reaudit found no improvement in results.
Conclusions:
Simple steps such as turning off computers and air-conditioning units can produce very significant financial and environmental savings. Radiology can lead the way in making hospitals more energy efficient.
More Related Videos
Related Concept Videos
Electrical Energy
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiation: Applications
The average...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...