Elevated ammonia concentrations: potential for pre-analytical and analytical contributing factors
Ibrahim A Hashim1, Jennifer A Cuthbert2
1Department of Pathology, University of Texas Southwestern Medical Center, Dallas, TX 75390, USA; Parkland Memorial Hospital, Parkland Health and Hospital System, Dallas, TX 75235, USA.
Background:
No study has explored the separate contributions of pre-analytical and analytical factors to hyperammonemia.
Methods:
Laboratory information systems were queried for tests of ammonia concentrations over a 12 month period. Pre-analytic (collection to laboratory receipt) and analytic (laboratory receipt to result) elapsed times were determined.
Results:
Under routine conditions for 3626 tests, normal and elevated results were similarly distributed if the time from venipuncture to result was <120 min. Delays, during analysis performance and in transportation to the laboratory, potentially contributed to hyperammonemia in a small number of samples (n=96, 2.7%). Similar results were obtained from a second hospital with a separate laboratory.
Conclusions:
Delays, in either transportation to the laboratory after collection or before completion of analysis, have the potential to elevate ammonia concentrations and may cause pseudo-hyperammonemia. Unexpectedly elevated ammonia concentrations need to be evaluated for errors in sampling handling.
More Related Videos
12:47Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Related Concept Videos
Urea Cycle
Hepatic Encephalopathy
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
EDTA: Auxiliary Complexing Reagents
Preparation of Amines: Alkylation of Ammonia and Amines
Each alkylation step makes the nitrogen center more nucleophilic, which triggers successive alkylations until a quaternary ammonium salt is formed. Considering...