Matthias Hammon1, Bernd Kunz2, Veronika Dinzl1
1Department of Radiology, University Hospital Erlangen, Erlangen, Bavaria, Germany.
Background:
To prove effectiveness of wrapping tablet computers in order to reduce microbiological contamination and to evaluate whether a plastic bag-covered tablet leads to impaired user satisfaction or touchscreen functionality.
Materials And Methods:
Within a period of 11 days 115 patients were provided with a tablet computer while waiting for their magnetic resonance imaging examination. Every day the contamination of the surface of the tablet was determined before the first and after the final use. Before the device was handed over to a patient, it was enclosed in a customized single-use plastic bag, which was analyzed for bacterial contamination after each use. A questionnaire was applied to determine whether the plastic bag impairs the user satisfaction and the functionality of the touchscreen.
Results:
Following the use by patients the outside of the plastic bags was found to be contaminated with various bacteria (657.5 ± 368.5 colony forming units/day); some of them were potentially pathogenic. In contrast, the plastic bag covered surface of the tablet was significantly less contaminated (1.7 ± 1.9 colony forming units/day). Likewise, unused plastic bags did not show any contamination. 11% of the patients reported problems with the functionality of the touchscreen. These patients admitted that they had never used a tablet or a smartphone before.
Conclusions:
Tablets get severely contaminated during usage in a clinical setting. Wrapping with a customized single-use plastic bag significantly reduces microbiological contamination of the device, protects patients from the acquisition of potentially pathogenic bacteria and hardly impairs the user satisfaction and the functionality of the touchscreen.
Related Concept Videos
Hand hygiene
Hand washing...
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated....
Handwashing III: During the Procedure and Post-Procedure Steps
Handwashing II: Pre-procedure and Initial Procedure Steps
The hand washing procedure itself includes the following steps. First, cover cuts, if any, on hands with a waterproof dressing. Cuts and abrasions can become contaminated with bacteria hindering the ability to clean the area thoroughly. In addition, repeated hand washing can worsen an injury. The nails must be short and clean, without nail...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Design Example: Resistive Touchscreen
When a user touches the screen, the two layers make contact at a specific point known as the touchpoint. This contact reduces the resistance between...