Related Experiment Video
Updated: Apr 24, 2026

Avocado Sample Preparation Using the QuEChERS Method with Ammonium Formate for Pesticide Analysis
Published on: December 8, 2023
[Determination of chloropicrin in air of workplace by gas chromatography with absorption solution collection]
Qitao Tan1, Huaisheng Bai1, Mingjian Gao1
1Zibo Hospital for Occupational Diseases Control and Treatment, Shandong 255000, China.
Objective:
To establish a method for determination of chloropicrin in the air of workplace by gas chromatography with liquid absorption collection.
Methods:
Chloropicrin in sample air was collected by absolute ethyl alcohol in a porous glass plate absorption tube. Following direct sample loading, samples were separated by gas chromatography and detected with a micro-electron capture detector. The present method was compared with the national standard method.
Results:
The range of linearity was 0-3.0 µg/ml, and the regression equation was y = 3488.80x-57.84, with a correlation coefficient of 0.999 4. The detection limit was 0.003 µg/ml, with a minimum detectable concentration of 0.005 mg/m³ based on a 3.0 L sampling volume. The average recovery rate was 98.9%∼103.2%. The within-run precision was 0.96%-2.12%, and the between-run precision was 2.75%-4.59%. The sampling efficiency was 98.6%. Samples in porous glass plate absorption tube could be stored at 4 °C for at least 5 days. The result ratio of this method to the national standard method was 97.7%-108.0% .
Conclusion:
The present method meets the requirements of "Guide for establishing occupational health standards-Part 4: Determination methods of air chemicals in workplace" (GBZ/T 210.4-2008), and is feasible for determination of chloropicrin in the air of workplace.
More Related Videos
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
07:32Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...