A single center experience with retrievable IVC filters.
Anas Renno1, Faisal Khateeb1, Viviane Kazan1
1Division of Vascular and Endovascular Surgery, University of Toledo Medical Center, Toledo, OH, USA.
This study evaluated retrievable inferior vena cava (IVC) filters, finding successful retrieval rates with minimal complications. Many IVC filters were permanently placed, and filter tilting was a common cause of retrieval failure.
Area of Science:
- Vascular Surgery
- Interventional Radiology
- Medical Device Evaluation
Background:
- Retrievable inferior vena cava (IVC) filters are used to prevent pulmonary embolism.
- Assessing the performance and retrieval rates of these devices is crucial for patient safety and optimal treatment.
- A structured follow-up program is essential for managing IVC filter retrieval.
Purpose of the Study:
- To evaluate the retrievable IVC filters used at our institution.
- To assess the success rates and complications associated with IVC filter retrieval.
- To analyze factors influencing filter retrieval and identify common reasons for failure.
Main Methods:
- Retrospective cohort study analyzing medical records of patients with retrievable IVC filters.
- Inclusion criteria: filter insertions between July 2007 and August 2011.
- Data analysis included demographics, indications, complications, retrieval rates, and filter brands; statistical analysis used Chi-square and t-tests.
Main Results:
- A total of 484 patients were reviewed; 258 had complete records. Over half (52.7%) of IVC filters were ultimately used as permanent.
- Retrieval was attempted in 73 patients (28.2%), with a 94.5% success rate. Filter tilting was the primary cause of retrieval failure.
- Bard filters were most commonly used as retrievable (80.9%), and retrieval success rates for Celect and Tulip filters were significantly lower than for Bard.
Conclusions:
- Retrievable IVC filters can be successfully retrieved with low complication rates.
- A significant proportion of IVC filters are converted to permanent use.
- Filter tilting is a key factor contributing to failed IVC filter retrieval.