A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Evaluation of a brachial cuff and suprasystolic waveform algorithm method to noninvasively derive central blood
Benedict T Costello1, Martin G Schultz2, J Andrew Black1
1Royal Hobart Hospital, Hobart Australia.
Background:
Central blood pressure (BP) can be estimated noninvasively by analyzing brachial artery waveforms. In this study, our aim was to assess the validity of a brachial cuff-based (suprasystolic) technique for estimating central BP (CBPestimated) by comparison with invasive aortic BP (CBPinvasive).
Methods:
Eighty-four simultaneous CBPestimated (Pulsecor R7.0) and CBPinvasive measures were recorded in 47 patients (aged 63±10 years, 62% male) undergoing coronary angiography. Measures were captured at baseline and acutely following intravenous glyceryl trinitrate (GTN; 100-200 μg). Mean CBPinvasive systolic BP (SBP) and diastolic BP (DBP) were compared with CBPestimated SBP and DBP calibrated with brachial SBP and DBP recorded from the Pulsecor device. To test validity of the central BP algorithm, measures of CBPestimated SBP were also compared with CBPinvasive SBP following recalibration with invasive mean arterial pressure (MAP) and DBP.
Results:
At baseline, mean difference ± standard deviation between CBPestimated SBP and CBPinvasive SBP was -7±9mm Hg (intraclass correlation coefficient (ICC) = 0.86; P < 0.001) with similar underestimation post-GTN (-6±9mm Hg; ICC = 0.90; P < 0.001). Recalibration of CBPestimated SBP with invasive MAP and DBP resulted in closer mean difference to CBPinvasive SBP (-2±7mm Hg; ICC = 0.95; P < 0.001) at baseline but not post-GTN (-6±7mm Hg; ICC = 0.95; P < 0.001).
Conclusions:
The Pulsecor algorithm to derive central BP has acceptable validity when calibrated with invasive pressures. However, accuracy is compromised when noninvasive brachial cuff BP calibration is used.
More Related Videos
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: