Brain tumours at 7T MRI compared to 3T--contrast effect after half and full standard contrast agent dose: initial
Iris-Melanie Noebauer-Huhmann1, P Szomolanyi, C Kronnerwetter
1High Field MR Centre, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, 1090, Vienna, Austria, iris.noebauer@medunwien.ac.at.
Objectives:
To compare the contrast agent effect of a full dose and half the dose of gadobenate dimeglumine in brain tumours at 7 Tesla (7 T) MR versus 3 Tesla (3T).
Methods:
Ten patients with primary brain tumours or metastases were examined. Signal intensities were assessed in the lesion and normal brain. Tumour-to-brain contrast and lesion enhancement were calculated. Additionally, two independent readers subjectively graded the image quality and artefacts.
Results:
The enhanced mean tumour-to-brain contrast and lesion enhancement were significantly higher at 7 T than at 3T for both half the dose (91.8 ± 45.8 vs. 43.9 ± 25.3 [p = 0.010], 128.1 ± 53.7 vs. 75.5 ± 32.4 [p = 0.004]) and the full dose (129.2 ± 50.9 vs. 66.6 ± 33.1 [p = 0.002], 165.4 ± 54.2 vs. 102.6 ± 45.4 [p = 0.004]). Differences between dosages at each field strength were also significant. Lesion enhancement was higher with half the dose at 7 T than with the full dose at 3T (p = .037), while the tumour-to-brain contrast was not significantly different. Subjectively, contrast enhancement, visibility, and lesion delineation were better at 7 T and with the full dose. All parameters were rated as good, at the least.
Conclusion:
Half the routine contrast agent dose at 7 T provided higher lesion enhancement than the full dose at 3T which indicates the possibility of dose reduction at 7 T.
Key Points:
• The contrast effect of gadobenate dimeglumine was assessed at 7 T and 3T. • In brain tumours, contrast effect was higher at 7 T than at 3T. • Tumour-to-brain contrast at 7 T half dose and 3T full dose were comparable. • 7 T half dose lesion enhancement was higher than 3T full dose enhancement. • Our results indicate the possibility of contrast agent dose reduction at 7 T.