3 Tesla magnetic resonance imaging noise in standard head and neck sequence does not cause temporary threshold shift
Elizabeth Yenn Lynn Lim1,2, Ing Ping Tang3, Mohammadreza Peyman4
1ORL Department, Faculty of Medicine, University Malaya, Kuala Lumpur, Malaysia. yennlynn@yahoo.com.
Abstract:
High acoustic noise level is one of the unavoidable side effects of 3 T magnetic resonance imaging (MRI). A case of hearing loss after 3 T MRI has been reported in this institution and hence this study. The objective of this study was to determine whether temporary threshold shift (TTS) in high frequency hearing occurs in patients undergoing 3 T MRI scans of the head and neck. A total of 35 patients undergoing head and neck 3 T MRI for various clinical indications were tested with pure tone audiometry in different frequencies including high frequencies, before and after the MRI scan. Any threshold change from the recorded baseline of 10 dB was considered significant. All patients were fitted with foamed 3 M earplugs before the procedure following the safety guidelines for 3 T MRI. The mean time for MRI procedure was 1,672 s (range 1,040-2,810). The noise dose received by each patient amounted to an average of 3,906.29% (1,415-9,170%). The noise dose was derived from a normograph used by Occupational Noise Surveys. This was calculated using the nomograph of L eq, L EX, noise dose and time. There was no statistically significant difference between the hearing threshold before and after the MRI procedures for all the frequencies (paired t test, P > 0.05). For patients using 3 M foamed earplugs, noise level generated by 3 T MRI during routine clinical sequence did not cause any TTS in high frequency hearing.
More Related Videos
07:13Modified Experimental Conditions for Noise-Induced Hearing Loss in Mice and Assessment of Hearing Function and Outer Hair Cell Damage
Published on: February 10, 2023
08:57Measuring the Influence of Magnetic Vestibular Stimulation on Nystagmus, Self-Motion Perception, and Cognitive Performance in a 7T MRT
Published on: March 3, 2023
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
NMR Spectrometers: Resolution and Error Correction
Atomic Nuclei: Magnetic Resonance
π Electron Effects on Chemical Shift: Overview