Femoral vein thrombosis following percutaneous placement of the Greenfield filter
K D Tobin1, S O Pais, C B Austin
1Department of Diagnostic Radiology, University of Maryland Medical System/Hospital, Baltimore 21201.
Abstract:
Following percutaneous insertion of the Greenfield inferior vena cava filter via the femoral route, patients were evaluated for femoral vein thrombosis by real-time ultrasound or Duplex sonography. In 23 patients (24 veins) the femoral vein was dilated with an 8-mm (24 F) angioplasty balloon catheter. Eight of 24 common femoral veins were thrombosed (33%). These findings indicate that thrombosis is a significant complication of percutaneous filter insertion. Dilation was performed in 20 patients (22 veins) with a 9-mm (27 F) balloon catheter; only three veins were thrombosed (14%). Over-dilation of the femoral vein to 27 F to accommodate the 24 F sheath/dilator set may decrease the incidence of thrombosis.