Over-response of synthetic microDiamond detectors in small radiation fields
Anna Ralston1, Madelaine Tyler, Paul Liu
1Chris O'Brien Lifehouse Cancer Centre, Sydney, Australia.
Abstract:
The recently commercialized PTW microDiamond detector (T60019) has been designed for use in small radiation fields. Here we report on the measurement of relative output ratios for small fields using five microDiamond detectors. All of the microDiamond detectors over-responded in fields smaller than 20 mm, by up to 9.3% for a 4 mm field. The over-response was independent of accelerator type and choice of collimation. The over-response was slightly larger than that observed in silicon diodes. Since all five microDiamond detectors showed the same over-response the corrections presented here should be transferable to other examples of the microDiamond detector, provided that the detector meets the manufacturing specifications and the beam characteristics are comparable.
Related Concept Videos
Determination of Crystal Structures
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Biological Effects of Radiation
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...