Reporting standards for Bland-Altman agreement analysis in laboratory research: a cross-sectional survey of current
Viswas Chhapola1, Sandeep Kumar Kanwal2, Rekha Brar3
1Department of Paediatrics, Division of Paediatric Intensive Care, Lady Hardinge Medical College and Associated Kalawati Saran Children's Hospital, New Delhi, India viswaschhapola@gmail.com.
Background And Objectives:
To carry out a cross-sectional survey of the medical literature on laboratory research papers published later than 2012 and available in the common search engines (PubMed, Google Scholar) on the quality of statistical reporting of method comparison studies using Bland-Altman (B-A) analysis.
Methods:
Fifty clinical studies were identified which had undertaken method comparison of laboratory analytes using B-A. The reporting of B-A was evaluated using a predesigned checklist with following six items: (1) correct representation of x-axis on B-A plot, (2) representation and correct definition of limits of agreement (LOA), (3) reporting of confidence interval (CI) of LOA, (4) comparison of LOA with a priori defined clinical criteria, (5) evaluation of the pattern of the relationship between difference (y-axis) and average (x-axis) and (6) measures of repeatability.
Results And Interpretation:
The x-axis and LOA were presented correctly in 94%, comparison with a priori clinical criteria in 74%, CI reporting in 6%, evaluation of pattern in 28% and repeatability assessment in 38% of studies.
Conclusions:
There is incomplete reporting of B-A in published clinical studies. Despite its simplicity, B-A appears not to be completely understood by researchers, reviewers and editors of journals. There appear to be differences in the reporting of B-A between laboratory medicine journals and other clinical journals. A uniform reporting of B-A method will enhance the generalizability of results.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Significance Testing: Overview
Data Validation
Key parameters for method validation include:
Bioequivalence Data: Statistical Interpretation
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...