Related Experiment Video
Updated: Apr 24, 2026

A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
Osmol gap method for the detection of diethylene glycol in human serum
Michael G Holland1, Jamie Nelsen1, Thomas G Rosano1
1Department of Emergency Medicine, SUNY Upstate Medical University/ Upstate New York Poison Center, Syracuse, New York/ Department of Occupational Medicine, Glens Falls Hospital, Glens Falls, New York, U.S.A (Holland MG, Nelsen J); Department of Pathology and Laboratory Medicine, Albany Medical Center Hospital and College, Albany, New York, U.S.A (Rosano TG).
Background:
Measurement of the osmol gap (OG) is a technique that is used frequently in toxic alcohol poisonings (ethylene glycol (EG) and methanol) as a rapid means to estimate exposure, and can be performed in virtually all hospital laboratories. The value of the OG has not been previously evaluated for diethylene glycol (DEG) exposures. The primary objective of this study was to evaluate the utility of the OG in estimating DEG serum concentrations using the most common formula that is currently used for estimating methanol, ethanol, and ethylene glycol concentrations.
Methods:
This was a controlled laboratory investigation using serum samples individually spiked with a known quantity of toxic alcohol compared to no toxic alcohol. Test samples were spiked with ethanol, DEG, EG, and methanol. Serum chemistries and osmolality and osmolarity were determined, and the OG was determined for each specimen.
Results:
The percent error of estimating DEG concentrations of 26.3% was similar to the mean percent error for estimating other alcohol concentrations, 30.5%±5.6% (P>0.05, 95% confidence interval 16.7%-44.3%).
Conclusions:
The severity of metabolic effects associated with DEG and the need to appropriately determine rescue treatments mandate early detection of significant exposures for effective triage and patient management. Our results indicate that the percent error of the osmol gap method for estimating DEG concentration is similar to that of other toxic alcohols; this simple technique could be a valuable clinical tool, since quantitative DEG analysis is rarely available.
Related Concept Videos
Freezing Point Depression and Boiling Point Elevation
The boiling point of a liquid is the temperature at which its vapor pressure is equal to ambient atmospheric pressure. Since the vapor pressure of a solution is lowered due to the presence of nonvolatile solutes, it stands to reason that the solution’s boiling point will subsequently be increased. Vapor pressure increases with temperature, and so a solution will require a higher temperature than will pure solvent to achieve any given vapor pressure, including one...