Related Experiment Video
Updated: Apr 24, 2026

A New Straightforward Method for Lipophilicity logP Measurement using 19F NMR Spectroscopy
Published on: January 30, 2019
Determination of the purity of pharmaceutical reference materials by 1H NMR using the standardless PULCON methodology
Yulia B Monakhova1, Matthias Kohl-Himmelseher2, Thomas Kuballa2
1Chemisches und Veterinäruntersuchungsamt (CVUA) Karlsruhe, Weissenburger Strasse 3, 76187 Karlsruhe, Germany; Bruker Biospin GmbH, Silberstreifen, 76287 Rheinstetten, Germany; Institute of Chemistry, Saratov State University, Astrakhanskaya Street 83, 410012 Saratov, Russia.
Abstract:
A fast and reliable nuclear magnetic resonance spectroscopic method for quantitative determination (qNMR) of targeted molecules in reference materials has been established using the ERETIC2 methodology (electronic reference to access in vivo concentrations) based on the PULCON principle (pulse length based concentration determination). The developed approach was validated for the analysis of pharmaceutical samples in the context of official medicines control, including ibandronic acid, amantadine, ambroxol and lercanidipine. The PULCON recoveries were above 94.3% and coefficients of variation (CVs) obtained by quantification of different targeted resonances ranged between 0.7% and 2.8%, demonstrating that the qNMR method is a precise tool for rapid quantification (approximately 15min) of reference materials and medicinal products. Generally, the values were within specification (certified values) provided by the manufactures. The results were in agreement with NMR quantification using an internal standard and validated reference HPLC analysis. The PULCON method was found to be a practical alternative with competitive precision and accuracy to the classical internal reference method and it proved to be applicable to different solvent conditions. The method can be recommended for routine use in medicines control laboratories, especially when the availability and costs of reference compounds are problematic.
More Related Videos
09:24Application and Methodology of the Non-destructive 19F Time-domain NMR Technique to Measure the Content in Fluorine-containing Drug Products
Published on: August 22, 2017
05:05Quantification of Polybutylene Adipate Terephthalate-based Micro- and Nano-plastics from Soil Using Proton Nuclear Magnetic Resonance Spectroscopy
Published on: June 6, 2025
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Other Nuclides: 31P, 19F, 15N NMR
While fluorine-19 and phosphorous-31 have high natural abundances (100%) and positive gyromagnetic ratios, nitrogen-15 has a low natural abundance and a negative gyromagnetic ratio. However, nitrogen-15 is still preferred over nitrogen-14 (which has a...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
High-Resolution Mass Spectrometry (HRMS)
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR