Related Experiment Video
Updated: Apr 24, 2026

A New Murine Model of Endovascular Aortic Aneurysm Repair
Published on: July 7, 2013
External aortic wrap for repair of type 1 endoleak
Anastasia Dean1, Swee Leong Yap1, Venu Bhamidipaty1
1Department of Vascular Surgery, Western Hospital, Gordon Street, Footscray, VIC 3011, Australia.
Introduction:
Type 1 endoleak is a rare complication after endovascular abdominal aortic aneurysm repair (EVAR) with a reported frequency up to 2.88%. It is a major risk factor for aneurysmal enlargement and rupture.
Presentation Of Case:
We present a case of a 68 year old gentleman who was found to have a proximal type 1 endoleak with loss of graft wall apposition on routine surveillance imaging post-EVAR. An initial attempt at endovascular repair was unsuccessful. Given the patient's multiple medical co-morbidities, which precluded the possibility of conventional graft explantation and open repair, we performed a novel surgical technique which did not require aortic cross-clamping. A double-layered Dacron wrap was secured around the infra-renal aorta with Prolene sutures, effectively hoisting the posterior bulge to allow wall to graft apposition and excluding the endoleak. Post-operative CT angiogram showed resolution of the endoleak and a stable sac size.
Discussion:
Several anatomical factors need to be considered when this technique is proposed including aortic neck angulation, position of lumbar arteries and peri-aortic venous anatomy. While an external wrap technique has been investigated sporadically for vascular aneurysms, to our knowledge there is only one similar case in the literature.
Conclusion:
Provided certain anatomical features are present, an external aortic wrap is a useful and successful option to manage type 1 endoleak in high-risk patients who are unsuitable for aortic clamping.