Middle cerebral artery blood flows by combining TCD velocities and MRA diameters: in vitro and in vivo validations
K A Yonan1, E R Greene2, J M Sharrar3
1Department of Biology and Chemistry and Department of Computer of Mathematical Sciences, New Mexico Highlands University, Las Vegas, New Mexico.
Abstract:
Non-invasive transcranial Doppler (TCD) is widely used for blood velocity (BV, cm/sec) measurements in the human middle cerebral artery (MCA). MCABV measurements are accepted as linear with MCA blood flow (MCABF). Magnetic resonance angiography (MRA) provides measurements of MCA lumen diameters that can be combined with TCD MCABV to calculate MCABF (mL/min). We tested the precision and accuracy of this method against a flow phantom and in vivo proximal internal carotid artery blood flow (ICABF). In vitro precision (repeated measures) and accuracy (vs. time collection) gave correlations coefficients of 0.97 and 0.98, respectively (both p < 0.05). In vivo precision (repeated measures) and accuracy (vs. ICABF) gave correlation coefficients of 0.90 (left and right), 0.94 (left) and 0.93 (right) (all p < 0.05). Bilateral MCABF in 35 adults were similar (left, 168 ± 72 mL/min; right, 180 ± 69 mL/min; p > 0.05). Results suggest that blood velocity by TCD and lumen diameter by MRA can be combined to estimate absolute values of MCABF.
More Related Videos
09:41Functional Transcranial Doppler Ultrasound for Monitoring Cerebral Blood Flow
Published on: March 15, 2021
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025