A comparison of right ventricular non-apical defibrillator lead position with traditional right ventricular apical
Gerald C Kaye1, Lim K Eng2, Benjamin J Hunt2
1University of Queensland, Brisbane, Queensland, Australia 4102.
Background:
Implantable cardioverter defibrillator (ICD) leads have traditionally been placed at the right ventricular apex (RVA). An important minority of patients with an ICD may develop a future requirement for bradycardia support. Pacing from the RVA may be detrimental, promoting heart failure and mortality. Increasingly non-apical right ventricular (RVNA) lead positions have been suggested as an alternative pacing site.
Methods:
A retrospective review of 512 patients who received an ICD at our institution between 1999 and 2011 was conducted. A comparison of lead performance characteristics was performed between RVNA sites and those at RVA. Data were collated from chart review and the pacing database.
Results:
The mean follow-up period in the RVNA cohort was 40.4±25.9 months and in the RVA cohort it was 38±31.8 months (p=0.43). The RVNA cohort consisted of 144 leads and 368 leads in the RVA cohort. The groups had similar baseline clinical characteristics. No significant difference was detected in the proportion of patients receiving an appropriate ICD defibrillation (RVNA 10.4% vs. RVA 16.8%; p=0.07), inappropriate defibrillation (RVNA 7.6% vs. RVA 7.6%; p=0.99) or an unsuccessful defibrillation (RVNA 0% vs. RVA 1.7%; p=0.12). There was also no significant difference in the proportion of patients receiving successful anti-tachycardia pacing (ATP) (RVNA 13.2% vs. RVA 17.4%; p=0.49) or failed ATP (RVNA 2.7% vs. RVA 4.1%; p=0.25). There was no significant difference in lead impedance (p=0.99), sensing (p=0.59) and pacing threshold (p=0.34).
Conclusion:
In this large retrospective study, RVNA ICD lead had similar stability and therapy efficacy compared to the traditional RVA position. This potentially has important implications for the suitability of RVNA as an alternative site for ICD leads.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Cardiac Catheterization II: Right Heart Catheterization
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation