A comparison of micro-CT and dental CT in assessing cortical bone morphology and trabecular bone microarchitecture
Jui-Ting Hsu1, Ying-Ju Chen2, Jung-Ting Ho1
1School of Dentistry, College of Medicine, China Medical University, Taichung, Taiwan.
Objective:
The objective of this study was to evaluate the relationship between the trabecular bone microarchitecture and cortical bone morphology by using micro-computed tomography (micro-CT) and dental cone-beam computed tomography (dental CT).
Materials And Methods:
Sixteen femurs and eight fifth lumbar vertebrae were collected from eight male Sprague Dawley rats. Four trabecular bone microarchitecture parameters related to the fifth lumbar vertebral body (percent bone volume [BV/TV], trabecular thickness [TbTh], trabecular separation [TbSp], and trabecular number [TbN]) were calculated using micro-CT. In addition, the volumetric cancellous bone grayscale value (vCanGrayscale) of the fifth lumbar vertebral body was measured using dental CT. Furthermore, four cortical bone morphology parameters of the femoral diaphysis (total cross-sectional area [TtAr], cortical area [CtAr], cortical bone area fraction [CtAr/TtAr], and cortical thickness [CtTh]) were calculated using both micro-CT and dental CT. Pearson analysis was conducted to calculate the correlation coefficients (r) of the micro-CT and dental CT measurements. Paired-sample t tests were used to compare the differences between the measurements of the four cortical bone morphology parameters obtained using micro-CT and dental CT.
Results:
High correlations between the vCanGrayscale measured using dental CT and the trabecular bone microarchitecture parameters (BV/TV [r = 0.84] and TbTh [r = 0.84]) measured using micro-CT were observed. The absolute value of the four cortical bone morphology parameters may be different between the dental CT and micro-CT approaches. However, high correlations (r ranged from 0.71 to 0.90) among these four cortical bone morphology parameters measured using the two approaches were obtained.
Conclusion:
We observed high correlations between the vCanGrayscale measured using dental CT and the trabecular bone microarchitecture parameters (BV/TV and TbTh) measured using micro-CT, in addition to high correlations between the cortical bone morphology measured using micro-CT and dental CT. Further experiments are necessary to validate the use of dental CT on human bone.