Related Experiment Video
Updated: Apr 23, 2026

Measurement of Bioelectric Current with a Vibrating Probe
Published on: January 4, 2011
Experimental verification of depolarization effects in bioelectrical impedance measurement
Xiaoyan Chen1, Xinqiang Lv1, Meng Du1
1College of Electronic Information and Automation, Tianjin University of Science and Technology, Tianjin 300222, China.
Abstract:
The electrode polarization effects on bioelectrical impedance measurement at low-frequency cannot be ignored. In this paper, the bioelectrical data of mice livers are measured to specify the polarization effects on the bio-impedance measurement data. We firstly introduce the measurement system and methodology. Using the depolarization method, the corrected results are obtained. Besides, the specific effects of electrode polarization on bio-impedance measurement results are investigated using comparative analysis of the previous and posterior correction results from dielectric spectroscopy, Cole-Cole plot, conductivity and spectroscopy of dissipation tangent. Experimental results show that electrode polarization has a significant influence on the characteristic parameters of mouse liver tissues. To be specific, we see a low-frequency limit resistance R0 increase by 19.29%, a reactance peak XP increase by 8.50%, a low-frequency limit conductivity Kl decrease by 17.65% and a dissipation peak tangent decrease by 160%.