Related Experiment Video
Updated: Apr 23, 2026

Genotyping of Staphylococcus aureus by Ribosomal Spacer PCR RS-PCR
Published on: November 4, 2016
Evaluation of the BD Max MRSA XT assay for use with different swab types
Alexander H Dalpke1, Marjeta Hofko2, Christian Stock3
1Department of Infectious Diseases, Medical Microbiology and Hygiene, University Hospital Heidelberg, Heidelberg, Germany alexander.dalpke@med.uni-heidelberg.de.
Abstract:
We evaluated the performance of the BD Max MRSA XT assay for use with different swab types. The 90% detection rates (95% confidence intervals) were 387 (97 to 1,551), 877 (238 to 3,230), 986 (183 to 5,287), 1,292 (328 to 5,078), 2,400 (426 to 13,518), and 5,848 (622 to 55,021) CFU/swab for Liquid Stuart, Liquid Amies, dry, Amies Gel without charcoal, ESwab collection, and Amies gel with charcoal swabs (Becton Dickinson), respectively. Amies Gel without charcoal, ESwab collection, and Amies gel with charcoal swabs had a tendency to be less sensitive, but none of the differences was statistically significant.
More Related Videos
09:03Multiplex PCR Assay for Typing of Staphylococcal Cassette Chromosome Mec Types I to V in Methicillin-resistant Staphylococcus aureus
Published on: September 5, 2013
07:50Author Spotlight: Scalable Drug Screening Protocol for Efficient Discovery of M. abscessus Treatments
Published on: October 25, 2024