Quantitation of zygomatic complex symmetry using 3-dimensional computed tomography
1Resident, Department of Oral and Maxillofacial Surgery, Peking University School and Hospital of Stomatology, Beijing, China.
Purpose:
To improve the clinical evaluation of the symmetry of the zygomatic complex (ZMC), the authors developed a new method to measure the eminence and width of the ZMC using 3-dimensional (3D) computed tomography (CT). The accuracy and reproducibility of the ZMC symmetry evaluation using this method were investigated.
Materials And Methods:
Spiral CT was carried out in 50 volunteers with visually symmetric faces. Based on an exact craniofacial midsagittal plane, a 3D coordinate system was constructed and used to measure the eminence and width of the ZMC. Absolute differences between the bilateral eminences and widths were calculated as ΔE and ΔW, respectively. The ZMC eminence also was measured based on traditional methods using anatomic landmarks, and the bilateral absolute difference was calculated to serve as a control (ΔcE).
Results:
The ΔE value was lower than the ΔcE value (0.6 vs 1.75 mm; P < .001). None of the results for ΔE were larger than 2 mm, and 7 results (3.5%) for ΔW were larger than 2 mm. By comparison, 57 results (28.5%) for ΔcE were larger than 2 cm (P < .001). The intraclass correlation coefficient (ICC) indicated almost perfect agreement for inter- and intraexaminer reliabilities for ΔE (ICC, 0.82 to 0.86) and ΔW (ICC, 0.81 to 0.85) measurements, and moderate agreement for inter- and intraexaminer reliabilities was obtained for ΔcE (ICC, 0.61 to 0.68) measurements.
Conclusion:
This new method is practical and valuable in the clinical evaluation of ZMC symmetry.
More Related Videos
02:42Analysis of Craniomaxillofacial Malformations in Mice Using Three-dimensional Microcomputed Tomography
Published on: January 17, 2025
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...