Improved accuracy of pulmonary embolism computer-aided detection using iterative reconstruction compared with
Kian Lahiji1, Seth Kligerman, Jean Jeudy
11 All authors: Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland Medical Center, 22 S Greene St, Baltimore, MD 21201.
Iterative reconstruction (IR) reduces false positives in pulmonary embolism computer-aided detection (PE CAD) but lowers sensitivity. High IR levels decrease detection accuracy, so they should be avoided with PE CAD systems.
Area of Science:
- Radiology
- Medical Imaging
- Computer-Aided Detection
Background:
- Pulmonary embolism (PE) is a critical diagnosis.
- Computer-aided detection (CAD) systems aim to improve PE detection accuracy.
- Image reconstruction techniques can impact CAD performance.
Purpose of the Study:
- To evaluate if iterative reconstruction (IR) enhances the performance of a pulmonary embolism computer-aided detection (PE CAD) prototype.
- To compare PE CAD performance using filtered back projection (FBP) versus various levels of hybrid IR.
Main Methods:
- CT angiographic images from 40 PE-positive and 26 control cases were analyzed.
- Images were reconstructed using filtered back projection (FBP) and six levels of hybrid iterative reconstruction (HIR).
- A prototype PE CAD system assessed performance on a per-embolus and per-study basis for each reconstruction type.
Main Results:
- Hybrid IR significantly reduced false-positive marks from PE CAD compared to FBP.
- Specificity improved with higher HIR levels, but sensitivity (both per-embolus and per-study) decreased progressively.
- Overall accuracy was highest at HIR level 1, with higher levels showing reduced performance.
Conclusions:
- Iterative reconstruction (IR) effectively reduces false positives in PE CAD.
- However, high levels of IR lead to decreased sensitivity and overall accuracy.
- Avoid using very high IR levels concurrently with PE CAD to maintain optimal detection performance.