Related Experiment Video
Updated: Apr 23, 2026

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Biomechanical comparison of sagittal-parallel versus non-parallel pedicle screw placement
Mazda Farshad1, Nadja A Farshad-Amacker, Elias Bachmann
1Orthopaedic Surgery, Balgrist University Hospital, University of Zürich, Forchstrasse 340, 8008, Zürich, Switzerland, mazda.farshad@balgrist.ch.
Background:
While convergent placement of pedicle screws in the axial plane is known to be more advantageous biomechanically, surgeons intuitively aim toward a parallel placement of screws in the sagittal plane. It is however not clear whether parallel placement of screws in the sagittal plane is biomechanically superior to a non-parallel construct. The hypothesis of this study is that sagittal non-parallel pedicle screws do not have an inferior initial pull-out strength compared to parallel placed screws.
Methods:
The established lumbar calf spine model was used for determination of pull-out strength in parallel and non-parallel intersegmental pedicle screw constructs. Each of six lumbar calf spines (L1-L6) was divided into three levels: L1/L2, L3/L4 and L5/L6. Each segment was randomly instrumented with pedicle screws (6/45 mm) with either the standard technique of sagittal parallel or non-parallel screw placement, respectively, under fluoroscopic control. CT was used to verify the intrapedicular positioning of all screws. The maximum pull-out forces and type of failure were registered and compared between the groups.
Results:
The pull-out forces were 5,394 N (range 4,221 N to 8,342 N) for the sagittal non-parallel screws and 5,263 N (range 3,589 N to 7,554 N) for the sagittal-parallel screws (p = 0.838). Interlevel comparisons also showed no statistically significant differences between the groups with no relevant difference in failure mode.
Conclusion:
Non-parallel pedicle screws in the sagittal plane have at least equal initial fixation strength compared to parallel pedicle screws in the setting of the here performed cadaveric calf spine experiments.