High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Improvement in PET/CT image quality in overweight patients with PSF and TOF
Takafumi Taniguchi1, Go Akamatsu, Yukiko Kasahara
1Division of Medical Quantum Science, Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan.
Objective:
The aim of this study was to evaluate the effect of the point spread function (PSF) and time of flight (TOF) on PET/CT images of overweight patients in relation to the iteration number and the acquisition time.
Methods:
This study consisted of a phantom study and a clinical study. The NEMA IEC body phantom and a 40 cm diameter large phantom (LG phantom) simulating an overweight patient were used in this study. Both phantoms were filled with (18)F solution with a sphere to background ratio of 4:1. The PET data were reconstructed with the baseline ordered-subsets expectation maximization (OSEM) algorithm, with the OSEM + PSF model, with the OSEM + TOF model and with the OSEM + PSF + TOF model. The clinical study was a retrospective analysis of 66 patients who underwent (18)F-FDG PET/CT. The image quality was evaluated using the background variability (coefficient of variance, CVphantom and CVliver) and the contrast (CONTHOT and SNR).
Results:
In phantom study, the CVphantom of the LG phantom was higher than that of the NEMA phantom. The PSF decreased the CVphantom of the LG phantom to the NEMA phantom level. The TOF information accelerated the CVphantom plateau earlier. The best relationship between the CVphantom and the CONTHOT was observed for the OSEM + PSF + TOF. In clinical study, the combination of PSF and TOF decreased the CVliver for overweight patients to that for normal weight patients while it increased the SNR similarly between two patient groups.
Conclusion:
The combination of the PSF and TOF correction improved the image quality of the LG phantom and overweight patients.
More Related Videos
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography