Radiographic predictors of lead conductor fracture
Andrew D Krahn1, Josée Morissette2, Ryan Lahm2
1From the University of British Columbia, Vancouver, British Columbia, Canada (A.D.K.); Medtronic, Minneapolis, MN (J.M., R.L., T.H., R.M.); Medtronic, Santa Ana, CA (W.W.B.); Sunnybrook and Women's Hospital, Toronto, Ontario, Canada (E.C.); Centre Hospitalier Universitaire de Sherbrooke, Sherbrooke, Quebec, Canada (F.A.-P.); University of Toronto, Toronto, Ontario, Canada (D.C.); Southlake Regional Hospital, Newmarket, Ontario, Canada (A.V.); Queen's University, Kingston, Ontario, Canada (C.S.S.); Libin Cardiovascular Institute, Calgary, Alberta, Canada (D.V.E.); and University of Ottawa Heart Institute, Ottawa, Ontario, Canada (D.H.B.). akrahn@mail.ubc.ca.
Background:
Lead fracture is a limiting factor in high voltage lead durability. Fractures noted with the Medtronic Fidelis leads provide an opportunity to examine factors captured on implant chest x-ray that correlate with risk for lead conductor fracture. We evaluated contributory factors in a large population of fractures.
Methods And Results:
We conducted a retrospective case-control study at 8 Canadian centers that routinely capture anterior posterior and lateral chest x-rays within 2 weeks of implant. Cases were patients that experienced confirmed Medtronic Fidelis 6949 lead fracture based on standard definitions, matched one-to-one to controls for date of implant, sex, and age with normally functioning Fidelis leads from the same center. Select chart data and x-rays were collected for all patients. Radiographic measurements by ≥2 individuals per case/control were blinded to patient status. The data were analyzed using a time to failure multivariable Cox proportional hazards model with stratification for each matched pair. X-ray pairs from 111 fracture patients were compared with 111 controls (age 61.5±12.8 years, 75% male, 221 model 6949 leads). Six parameters included in the statistical analysis were significantly associated with risk of fracture, including slack/tortuosity measures, pulse generator and superior vena cava coil location, and angle of lead exit from the pocket.
Conclusions:
Pocket, intravascular and intracardiac lead characteristics on x-ray correlate with risk of lead conductor fracture. These observations may be useful to direct implant technique to optimize lead durability. Validation in larger populations and other lead models may inform the application of these results.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...