Related Experiment Video
Updated: Apr 23, 2026

Computational Reconstruction of Pancreatic Islets as a Tool for Structural and Functional Analysis
Published on: March 9, 2022
Design and in silico evaluation of an intraperitoneal-subcutaneous (IP-SC) artificial pancreas
Justin J Lee1,2, Eyal Dassau1,2, Howard Zisser1,2
1Department of Chemical Engineering, The University of California, Santa Barbara, CA 93106-5080, USA.
Abstract:
Prandial glucose regulation is a major challenge for the artificial pancreas using subcutaneous insulin (without a feedforward bolus) due to insulin's slow absorption-peak (50-60 min). Intraperitoneal insulin, with a fast absorption peak (20-25 min), has been suggested as an alternative to address these limitations. An artificial pancreas using intraperitoneal insulin was designed and evaluated on 100 in silico subjects and compared with two designs using subcutaneous insulin with and without a feedforward bolus, following the three meal (40-70 g-carbohydrates) evaluation protocol. The design using intraperitoneal insulin resulted in a significantly lower postprandial blood glucose peak (38 mg/dL) and longer time in the clinically accepted region (13%) compared to the design using subcutaneous insulin without a feedforward bolus and comparable results to a subcutaneous feedforward bolus design. This superior regulation with minimal user interaction may improve the quality of life for people with type 1 diabetes mellitus.