Variant parameter values-as defined by the Chicago Criteria-produced by ManoScan and a new system with Unisensor
S Kuribayashi1, K Iwakiri, A Kawada
1Department of Gastroenterology, Gunma University Hospital, Gunma, Japan; Department of Gastroenterology and Hepatology, National Hospital Organization Numata National Hospital, Gunma, Japan.
Background:
Recently reported normal values for esophageal motility obtained by high-resolution manometry (HRM) using a system with a Unisensor catheter were significantly different from those obtained by the ManoScan(®) , which could result in a wrong diagnosis. To clarify whether these differences were due to system or subject differences, we compared the manometric parameter values between ManoScan and a new system with a Unisensor catheter (Starlet) in the same subjects.
Methods:
A total of 103 volunteers without any symptoms related to esophageal motility disorders were recruited. Esophageal HRM was performed using both the ManoScan and the Starlet in all subjects. Data from the ManoScan were analyzed using ManoView, and data from the Starlet were analyzed by a program with e-sleeve function. Integrated relaxation pressure, distal contractile integral, contractile front velocity (CFV), intrabolus pressure, and distal latency were calculated by both analyzing programs, and the values of these parameters were compared between the two systems by a signed rank test.
Key Results:
Data from a total of 97 participants were analyzed. The values of all parameters, except CFV, measured by the Starlet were significantly higher than those obtained by the ManoScan (p < 0.01).
Conclusions & Inferences:
Both systems can measure esophageal motility appropriately; nevertheless, we confirmed that the two systems showed different values of the parameters defined by the Chicago criteria. These differences should be recognized to evaluate esophageal motility precisely.
More Related Videos
09:05Catheterization of the Carotid Artery and Jugular Vein to Perform Hemodynamic Measures, Infusions and Blood Sampling in a Conscious Rat Model
Published on: January 30, 2015
10:28Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization