Portal monitor characterization for internally and externally deposited radionuclides
Matthew Carey1, Adam Kryskow, Fred Straccia
1*University of Massachusetts, Lowell, Radiological Sciences, University Avenue, MA 01854.
Abstract:
In this evaluation, both the internal and external radionuclide detection efficiencies for a portal monitor were evaluated as a function of photon energy using an anthropomorphic phantom. Pass-through and static measurements were completed using 241Am, 57Co, 133Ba, 137Cs, 60Co, 109Cd, and 54Mn in various locations both external and internal to the phantom. Other parameters, such as single detector uniformity, total detector uniformity, background linearity, and activity linearity have been analyzed. It was found that the minimum detectable activity for internally deposited 137Cs in the abdomen was approximately ten times higher for pass-through versus static measurements. Additionally, it was found that the minimum detectable activity for 137Cs in the abdominal region for both internal and external pass-through scenarios are nearly equivalent. In general, if the expected radionuclide source term is primarily non-transuranic, the pass-through mode offers sufficient sensitivity to identify potential overexposures while providing much greater personnel throughput. However, minimum detectable committed effective doses for transuranics such as 241Am, show potential for personnel over exposure if the radionuclide mixture contains a significant fraction of transuranics. It is therefore recommended that nuclear facilities evaluate their radionuclide source term in order to bound potential personnel doses.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...