Related Experiment Video
Updated: Apr 23, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Catastrophic stem taper wear in ceramic on polyethylene bearing couple: a case report
Atsushi Kusaba1, Mariko Katsui2, Naoyuki Hakuta2
1Institute for Joint Replacement and Rheumatology, Ebina General Hospital, Kanagawa, Japan.
Abstract:
A severe abrasion occurred at the stem taper 13 years after total hip arthroplasty. The bearing couple was 28 mm ceramic on polyethylene. The ceramic head had almost no damage, whereas the stem taper had severe damage. An ectopic bone probably contributed to the abrasion. The ectopic bone had grown to restrict the normal motion in the bearing couple and all stress and torque seemed to concentrate in the junction until the junction obtained mobility instead of the original bearing couple.