[Dynamic characteristics of flow trainer for dpi and its clinical relevance]
Tetsuri Kondo1, Takanori Shimizu2, Toshimori Tanigaki3
1Department of Respiratory Medicine, Shonan Fujisawa Tokushukai Hospital.
Background:
Trainer devices are widely employed for flow instruction in the use of dry powder inhalers (DPI). However, their aerodynamic characteristics in actual use have yet to be investigated.
Methods:
We recorded inhalation flow signals and sounds produced from trainers for Diskus®, Symbicort®, Pulmicort® and Twisthaler® while five volunteers inhaled from the trainers with various inhalation patterns.
Results:
Inhalation flow was classified into four patterns; the best, trapezoid, delayed peak and others. All the trainers exhibited flow-thresholds with appropriate precision for instruction. Sound intensity from the Diskus® trainer was almost proportional to inhaled flow rate, and it may be useful for flow pattern estimation. In other trainers, when flow exceeded some thresholds, sounds abruptly developed and continued with high intensity. Thus, they may be convenient for recognizing appropriate flow rates. In all trainers, when the subjects inhaled rapidly and forcefully, sound developed at 0.19-0.24 s after the onset of inhalation. Thus, with this flow pattern, trainers may indicate a flow rate approaching the peak of drug dispersion from the DPI. When the subject inhaled less rapidly, the threshold for sound development decreased by 10%.
Conclusion:
The instructor in DPI use should be aware of the aerodynamic characteristics of each individual trainer. Rapid inhalation should also be encouraged.
More Related Videos
04:59Author Spotlight: Developing a Disposable Dosator for Preclinical Testing of Dry Powder Inhalers in Small Animal Models
Published on: August 18, 2023
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Inhaled Medications
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Pipe Flowrate Measurement: Problem Solving