A cryogen-free ultralow-field superconducting quantum interference device magnetic resonance imaging system
Byeong Ho Eom1, Konstantin Penanen1, Inseob Hahn1
1Jet Propulsion Laboratory, California Institute of Technology, Pasadena, California 91109, USA.
Abstract:
Magnetic resonance imaging (MRI) at microtesla fields using superconducting quantum interference device (SQUID) detection has previously been demonstrated, and advantages have been noted. Although the ultralow-field SQUID MRI technique would not need the heavy superconducting magnet of conventional MRI systems, liquid helium required to cool the low-temperature detector still places a significant burden on its operation. We have built a prototype cryocooler-based SQUID MRI system that does not require a cryogen. The SQUID detector and the superconducting gradiometer were cooled down to 3.7 K and 4.3 K, respectively. We describe the prototype design, characterization, a phantom image, and areas of further improvements needed to bring the imaging performance to parity with conventional MRI systems.