Simple and robust ICP-MS method for simultaneous determination of serum Co and Cr in routine clinical practice
Hyun-Jung Choi1, Seung-Jae Lim2, Youn-Soo Park2
1Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; Department of Laboratory Medicine, Chonnam National University Medical School and Chonnam National University Hospital, Gwanju, Korea.
Background:
Chromium (Cr) and cobalt (Co) metal ions released from orthopedic prostheses, such as metal-on-metal (MoM) bearings, may be related to wear of the bearing surface. These ions may serve as indicators of the in vivo performance of MoM bearing surfaces. We developed a simple and sensitive assay suitable for the simultaneous measurement of serum Cr and Co using inductively coupled plasma-mass spectrometry (ICP-MS) in clinical routine laboratory practice to evaluate analytical performance and clinical utility.
Methods:
We evaluated the linearity, precision, and lower limit of quantification (LOQ) of an ICP-MS method to determine serum Cr and Co concentration in accordance with U.S. Food and Drug Administration guidance for bioanalytical method validation. This method was used to determine the serum Cr and Co levels of 236 clinical samples from 74 patients with MoM hip prosthesis and 51 healthy controls.
Results:
The assay showed good linearity (0-20 μg/l, r(2)>0.999) and satisfactory accuracy for all tested concentrations of Cr and Co (%bias, -1.5% to 2.5%, -3.3% to 1.6%, respectively). The CVs for both metal ions were <10% for LOQ (0.02 μg/l Cr, 0.01 μg/l Co) and within 5% for other concentrations (intra- and inter-assay CV, 1.2-2.6 and 1.9-4.4% of Cr; 1.4-2.7 and 1.9-4.7% of Co). The serum Cr and Co concentrations (mean ± SD) were 0.60 ± 0.12 μg/l and 0.29 ± 0.15 μg/l in 51 healthy subjects. In 74 patients (185 serum samples) after HRA (median duration implanted, 48 months; range, 1-138 months), the concentrations of Cr and Co were 2.75 μg/l (range, 0.61-116.80 μg/l) and 1.49 μg/l (range, 0.12-127.80 μg/l), respectively.
Conclusions:
ICP-MS can simultaneously quantify serum Cr and Co levels with suitable analytical performance including a wide analytical range and very low LOQ. This assay will be very useful for evaluating the nutritional status of Cr and Co in healthy humans as well as monitoring the performance of orthopedic prostheses after hip resurfacing arthroplasty.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Precipitation and Co-precipitation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....