Related Experiment Video
Updated: Apr 23, 2026

Athymic Rat Model for Evaluation of Engineered Anterior Cruciate Ligament Grafts
Published on: March 26, 2015
Broken Bioabsorbable Tibial Interference Screw after Anterior Cruciate Ligament (ACL) Reconstruction using a
Me Deborah Huang1, Hc Andrew Tan1
1Department of Orthopaedic Surgery, Singapore General Hospital, Singapore.
Abstract:
When a patient presents with knee pain and locking after an anterior cruciate ligament (ACL) reconstruction, a new meniscal injury or an osteochondral loose body are usually considered for differential diagnosis. We present the case of a 22-year-old female with just these complaints 6 months after ACL reconstruction surgery. Magnetic resonance imaging (MRI) of the knee showed a broken screw tip which was later arthroscopically removed. At arthroscopy, an 11mm long broken bioabsorbable interference screw tip was found lying in the intercondylar notch; this resulted in a 0.5cm Outerbridge grade II chondral ulcer located at midpatella. Both menisci and cruciate ligaments were intact and no other loose bodies were found in the knee joint.
Key Words:
ACL reconstruction, bioabsorbable interference screw, chondral ulcer.