Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Bodily fluid analysis of non-serum samples using point-of-care testing with iSTAT and Piccolo analyzers versus a
William Londeree1, Konrad Davis1, Donald Helman1
1Department of Medicine, Tripler Army Medical Center, Honolulu, HI (WL).
Introduction:
Forward deployed military medical units can provide sophisticated medical care with limited resources. Point-of-Care Testing (POCT) may facilitate care and expedite diagnosis. This study assessed the accuracy of results for POCT for non-serum samples (pleural, peritoneal, and cerebrospinal fluid) using iSTAT and Piccolo hand-held devices compared with results obtained using a hospital chemistry analyzer.
Methods:
Pleural, peritoneal, and cerebrospinal fluids obtained during routine care were simultaneously analyzed on a Vitros 5600 automated clinical chemistry hospital analyzer, iSTAT, and Piccolo POCT devices.
Results:
POCT results were highly correlated with the Vitros 5600 for pleural fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG, glucose, albumin, and protein; and glucose for cerebrospinal fluid.
Conclusion:
POCT results for non-serum samples from pleural, peritoneal, and cerebrospinal fluid correlate with standard hospital chemistry analysis. The results of this study demonstrate potential for possible new diagnostic roles for POCT in resource-limited environments.