Electrically evoked compound action potentials artefact rejection by independent component analysis: procedure
Idrick Akhoun1, Colette McKay2, Wael El-Deredy3
1School of Psychological Sciences, Faculty of Medical and Human Sciences, The University of Manchester, Manchester, UK; Advanced Bionics European Research Centre, Feodor-Lynen-Strasse, 35, Hanover, Germany.
This study automates electrically-evoked compound action potential (ECAP) analysis using Independent Component Analysis (ICA). The automated ECAP-ICA method effectively removes artefacts, offering a less biased alternative for cochlear implant fitting.
Area of Science:
- Biomedical Engineering
- Neuroscience
- Signal Processing
Background:
- Independent Component Analysis (ICA) has proven effective in separating electrically-evoked compound action potentials (ECAPs) from noise and stimulation artefacts.
- Previous methods relied on manual artefact rejection, introducing potential biases.
Purpose of the Study:
- To develop and validate an automated artefact cancellation process for ECAP analysis using ICA.
- To provide a less biased alternative to traditional artefact rejection methods in cochlear implant research.
Main Methods:
- Raw ECAP data were recorded from intra-cochlear electrodes without initial artefact rejection.
- Amplifier-saturated recordings were discarded, and data from various stimulus conditions were concatenated.
- Independent Component Analysis (ICA) was applied with a restriction to 4 sources, enabling automatic classification of artefact, noise, and ECAP components.
Main Results:
- The automated ECAP-ICA procedure was tested on 8 cochlear implant users.
- Artefact and noise components were successfully identified and discarded, isolating the ECAP.
- The automated method accurately extracted ECAPs in 22 out of 26 cases when compared to the standard clinical forward masking paradigm.
Conclusions:
- The automated ECAP-ICA procedure offers a robust and unbiased method for ECAP extraction.
- This technique eliminates the need for complex buffer combinations required by traditional methods.
- ECAP-ICA holds significant clinical relevance, particularly as a component in automated ECAP-threshold detection for cochlear implant fitting software.