Related Experiment Video
Updated: Apr 23, 2026

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Infections and exposures: reported incidents associated with unsuccessful decontamination of reusable surgical
1Health Protection Scotland, Glasgow, UK.
Abstract:
Reusable surgical instruments provide a potential route for the transmission of pathogenic agents between patients in healthcare facilities. As such, the decontamination process between uses is a vital component in the prevention of healthcare-associated infections. This article reviews reported outbreaks and incidents associated with inappropriate, inadequate, or unsuccessful decontamination of surgical instruments, indicating potential pitfalls of decontamination practices worldwide. To the author's knowledge, this is the first review of surgical instrument decontamination failures. Databases of medical literature, Medline and Embase, were searched systematically. Articles detailing incidents associated with unsuccessful decontamination of surgical instruments were identified. Twenty-one articles were identified reporting incidents associated with failures in decontamination. A large proportion of incidents involved the attempted disinfection, rather than sterilization, of surgical instruments (43% of articles), counter to a number of national guidelines. Instruments used in eye surgery were most frequently reported to be associated with decontamination failures (29% of articles). Of the few articles detailing potential or confirmed pathogenic transmission, Pseudomonas aeruginosa and Mycobacterium spp. were most represented. One incident of possible variant Creutzfeldt-Jakob disease transmission was also identified. Limitations of analysing only published incidents mean that the likelihood of under-reporting (including reluctance to publish failure) must be considered. Despite these limitations, the small number of articles identified suggests a relatively low risk of cross-infection through reusable surgical instruments when cleaning/sterilization procedures are adhered to. The diverse nature of reported incidents also suggests that failures are not systemic.
Insights
Reusable surgical instruments can transmit pathogens, making proper decontamination crucial for preventing healthcare-associated infections. Failures in decontamination, though rare, highlight the need for adherence to sterilization guidelines to ensure patient safety.
Area of Science:
- Healthcare-associated infections
- Infection control
- Surgical instrument reprocessing
Background:
- Reusable surgical instruments pose a risk for pathogen transmission between patients.
- Effective decontamination is essential for preventing healthcare-associated infections.
Purpose of the Study:
- To review reported outbreaks and incidents linked to surgical instrument decontamination failures.
- To identify potential pitfalls in global decontamination practices.
Main Methods:
- Systematic search of Medline and Embase databases.
- Identification and analysis of articles detailing incidents of unsuccessful surgical instrument decontamination.
Main Results:
- Twenty-one articles reported decontamination failures.
- Disinfection, not sterilization, was attempted in 43% of incidents.
- Eye surgery instruments were most frequently implicated (29%).
- Pseudomonas aeruginosa and Mycobacterium spp. were common pathogens; possible variant Creutzfeldt-Jakob disease transmission noted.
Conclusions:
- While under-reporting is possible, the low number of identified incidents suggests a relatively low risk of cross-infection when procedures are followed.
- Failures in surgical instrument decontamination are diverse and not indicative of systemic issues.
More Related Videos
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
PPE Use in Healthcare Settings II: Doffing
Handwashing III: During the Procedure and Post-Procedure Steps