Influence of axial length on thickness measurements using spectral-domain optical coherence tomography
Tobias Röck1, Karl Ulrich Bartz-Schmidt1, Matthias Bramkamp2
1Centre for Ophthalmology, University of Tübingen, Tübingen, Germany.
Purpose:
The purpose of this study was to assess the influence of axial length on spectral-domain optical coherence tomography (SD-OCT) thickness measurements in patients with subretinal visual implants.
Methods:
Data from eight emmetropic pseudophakic eyes of eight patients with subretinal visual implants were analyzed retrospectively. These patients participated in the monocentric part of a multicenter trial. The axial length was measured in three short (<22.5 mm), three medium (22.51-25.50 mm), and two long (>25.52 mm) eyes. Using Heidelberg Spectralis, the known thickness of a subretinal implant microchip (70 μm) was measured on 15 images per eye with SD-OCT, using the software calipers.
Results:
The mean axial length was 20.8 ± 0.8 mm in short eyes, 23.3 ± 0.4 mm in medium eyes, and 26.3 ± 0.5 mm in long eyes. We found in short eyes, in medium eyes, and in long eyes a mean value of microchip thickness measurements from SD-OCT of 82.9 ± 1.4 μm, 70.5 ± 1.3 μm, and 64.2 ± 1.3 μm, respectively. The thickness measurements decreased in SD-OCT measurements with longer axial lengths significantly (P < 0.0001).
Conclusions:
Axial length influences SD-OCT thickness measurements. Our findings demonstrate accuracy of the scaling in SD-OCT thickness measurements in emmetropic medium eyes. Caution is recommended when comparing the measured values of short and long eyes with the normative database of the instrument. There is a need for larger sample-size studies to confirm our results. (ClinicalTrials.gov number, NCT01024803.).
More Related Videos
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
07:28Thinned-skull Cortical Window Technique for In Vivo Optical Coherence Tomography Imaging
Published on: November 19, 2012
Related Concept Videos
Thin-Walled Hollow Shafts
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...