Wear of highly crosslinked polyethylene acetabular components
Stuart A Callary1, Lucian B Solomon, Oksana T Holubowycz
1Department of Orthopaedics and Trauma, Royal Adelaide Hospital.
Acta Orthopaedica
|October 11, 2014
Summary
Highly crosslinked polyethylene (XLPE) acetabular components show low wear rates in total hip arthroplasty (THA) when measured by radiostereometric analysis (RSA). Standardizing reporting will improve implant performance evaluation.
Area of Science:
- Orthopedic surgery
- Biomaterials science
- Medical imaging analysis
Background:
- In vivo wear rates of highly crosslinked polyethylene (XLPE) acetabular components vary significantly across studies.
- Measurement techniques and inherent errors complicate accurate wear assessment in XLPE bearings.
- Radiostereometric analysis (RSA) offers the most sensitive method for in vivo wear measurement.
Purpose of the Study:
- To conduct a scoping review of studies measuring in vivo wear of XLPE acetabular components using RSA.
- To synthesize findings on XLPE wear rates in primary total hip arthroplasty (THA).
- To identify limitations in current RSA methodologies and reporting standards.
Main Methods:
- Systematic literature search of PubMed, Scopus, and Cochrane databases.
- Inclusion of studies utilizing RSA for wear measurement in primary THA with XLPE components.
- Analysis of 18 publications encompassing 12 primary THA cohorts.
Main Results:
- Reviewed studies reported mean/median proximal wear rates ranging from 0.00 to 0.06 mm/year.
- Variations in wear determination methods hindered direct study comparison.
- Differences in RSA techniques (radiograph positioning, reference markers) may limit future meta-analyses.
Conclusions:
- This review confirms low in vivo wear rates for XLPE acetabular components in THA, as evidenced by RSA.
- Standardization of RSA wear reporting is recommended for early detection of implant issues.
- Improved reporting will enhance understanding of patient and surgical factors influencing wear.