Related Experiment Video
Updated: Apr 22, 2026

Direct Analysis of Single Cells by Mass Spectrometry at Atmospheric Pressure
Published on: September 4, 2010
A fully-automated analyzer for determining haloacetic acid concentrations in drinking water
Christina M Henson1, Gary L Emmert1, Paul S Simone1
1Department of Chemistry, The University of Memphis, Memphis, TN 38152, USA.
Abstract:
A fully-automated, on-line, real-time analyzer has been developed for preconcentration and analysis of haloacetic acids (HAAs). Preconcentration of HAAs is achieved by sample acidification and solid phase extraction onto a hydrophobic polymeric resin using sequential injection analysis (SIA). The HAAs preconcentrate is then analyzed using post-column reaction-ion chromatography (PCR-IC), which is selective for HAAs. Systematic optimization of SIA preconcentration parameters are described followed by detailed method detection limit (MDL), accuracy, precision, and linearity studies. MDL values for the individual HAA9 species range from 0.4 to 0.9 μg L(-1). Side-by-side comparison studies of HAAs analysis in 14 real-world drinking water samples from Alabama, Arkansas, Kentucky, Minnesota, Missouri, Mississippi, New York, Pennsylvania and Tennessee are presented that compare the optimized SIA-PCR-IC to USEPA Method 552.3. Trace levels of HAAs detected in select samples are reported, and the bias values calculated between the two methods are typically less than 5 μg L(-1) for eight of the nine individual HAAs.
More Related Videos
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Related Concept Videos
Automated Microbial Diagnostics
Weak Acid Solutions
Titration in Nonaqueous Solvents
Water: A Bronsted-Lowry Acid and Base
Solution Composition During Acid/Base Titrations
The α0 and α1 values...
Acid-Catalyzed α-Halogenation of Aldehydes and Ketones
In the first step of the mechanism, the acid protonates the carbonyl oxygen resulting in a resonance-stabilized cation, which subsequently loses an α-hydrogen to form an enol tautomer. The C=C bond in an enol is highly nucleophilic because of the electron-donating nature of the –OH group. Consequently, the double bond attacks an electrophilic...