Related Experiment Video
Updated: Apr 22, 2026

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
An optically coupled sensor for the measurement of currents induced by MRI gradient fields into endocardial leads
Eugenio Mattei1, Federica Censi, Michele Triventi
1Department of Technology and Health, Italian National Institute of Health, Viale Regina Elena 299, 00161, Rome, Italy, eugenio.mattei@iss.it.
Object:
The gradient fields generated during magnetic resonance imaging (MRI) procedures have the potential to induce electrical current on implanted endocardial leads. Whether this current can result in undesired cardiac stimulation is unknown.
Materials And Methods:
This paper provides a detailed description of how to construct an optically coupled sensor for the measurement of gradient-field-induced currents into endocardial leads. The system is based on a microcontroller that works as analog-to-digital converter and sends the current signal acquired from the lead to an optical high-speed, light-emitting diode transmitter. A plastic fiber guides the light outside the MRI chamber to a photodiode receiver and then to an acquisition board connected to a PC laptop.
Results:
The performance of the system has been characterized in terms of power consumption (8 mA on average), sampling frequency (20.5 kHz), measurement range (-12.8 to 10.3 mA) and resolution (22.6 µA). Results inside a 3 T MRI scanner are also presented.
Conclusions:
The detailed description of the current sensor could permit more standardized study of MRI gradient current induction in pacemaker systems. Results show the potential of gradient currents to affect the pacemaker capability of triggering a heartbeat, by modifying the overall energy delivered by the stimulator.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Field Of A Current Loop
Magnetic Resonance Imaging
Magnetic Force Between Two Parallel Currents
The force exerted by the magnetic field due to the first conductor over a finite length of the second conductor is given as the product of the current in the second conductor and the vector product of the length vector along the current element and the field due to the first conductor. According to the...
Magnetic Force On Current-Carrying Wires: Example