A dual-use laryngoscope to facilitate apneic oxygenation
Thomas Mitterlechner1, Holger Herff2, Christian W Hammel1
1Department of Anesthesiology and Critical Care Medicine, Innsbruck Medical University, Innsbruck, Austria.
Background:
In preoxygenated patients, time until oxygen saturation drops can be extended by insufflating oxygen into their airways, thus oxygenating them apneically.
Objectives:
To compare different methods of apneic oxygenation.
Methods:
A noncommercial dual-use laryngoscope with an internal lumen in its blade was used to provide oxygen insufflation into a simulated laryngeal space during intubation. In this experimental study, oxygen insufflation via the dual-use laryngoscope was compared with no oxygen insufflation, with nasal oxygen insufflation, and with direct intratracheal oxygen insufflation. In a preoxygenated test lung of a manikin, oxygen percentage decrease was measured over a 20-min observation period for each method of oxygen application.
Results:
Oxygen percentage in the test lung dropped from 97% to 37 ± 1% in the control group (p < 0.001 compared to all other groups) and to 68 ± 1% in the nasal insufflation group (p < 0.001 compared to all other groups). Oxygen percentage remained over 90% in both the direct intratracheal insufflation group (96 ± 0%) and the laryngoscope blade insufflation group (94 ± 1%) (p < 0.01 between the latter two groups).
Conclusions:
Simulating apneic oxygenation in a preoxygenated manikin, deep laryngeal oxygen insufflation via the dual-use laryngoscope kept oxygen percentage in the test lung above 90%, and was more effective than oxygen insufflation via nasal prongs.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...