Related Experiment Video
Updated: Apr 22, 2026

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Biomechanical evaluation of the pedicle screw insertion depth effect on screw stability under cyclic loading and
Kristophe J Karami1, Laura E Buckenmeyer, Ata M Kiapour
1*Department of Neurosurgery, St John Providence Hospital and Medical Centers, Michigan State University, Southfield, MI †Department of Neurosurgery, Johns Hopkins University, Baltimore, MD ‡Engineering Center for Orthopaedic Research Excellence (ECORE) §Departments of Bioengineering and Orthopaedic Surgery, University of Toledo, Toledo, OH ∥Biomechanics & Injury Mitigation Systems, Research & Exploratory Development Department, The Johns Hopkins University Applied Physics Laboratory, Laurel, MD ¶Michigan Spine & Brain Surgeons, PLLC, Southfield, MI.
Study Design:
A biomechanical ex vivo study of the human lumbar spine.
Objective:
To evaluate the effects of transpedicular screw insertion depth on overall screw stability and pullout strength following cyclic loading in the osteoporotic lumbar spine.
Summary Of Background Data:
Although much is known about the clinical outcomes of spinal fusion, questions remain in our understanding of the biomechanical strength of lumbar pedicle screw fixation as it relates to screw sizing and placement. Biomechanical analyses examining ideal pedicle screw depth with current pedicle screw technology are limited. In the osteoporotic spine, optimized pedicle screw insertion depth may improve construct strength, decreasing the risk of loosening or pullout.
Methods:
A total of 100 pedicles from 10 osteoporotic lumbar spines were randomly instrumented with pedicle screws in mid-body, pericortical, and bicortical depths. Instrumented specimens underwent cyclic loading (5000 cycles of ±2 N m pure flexion moment) and subsequent pullout. Screw loosening, failure loads, and energy absorption were calculated.
Results:
Cyclic loading significantly (P<0.001) reduced screw-bone angular stiffness between prefatigue and postfatigue conditions by 25.6%±17.9% (mid-body), 20.8%±14.4% (pericortical), and 14.0%±13.0% (bicortical). Increased insertion depth resulted in lower levels of reduction in angular stiffness, which was only significant between mid-body and bicortical screws (P=0.009). Pullout force and energy of 583±306 N and 1.75±1.98 N m (mid-body), 713±321 N and 2.40±1.79 N m (pericortical), and 797±285 N and 2.97±2.33 N m (bicortical) were observed, respectively. Increased insertion depth resulted in higher magnitudes of both pullout force and energy, which was significant only for pullout force between mid-body and bicortical screws (P=0.005).
Conclusion:
Although increased screw depth led to increased fixation and decreased loosening, additional purchase of the stiff anterior cortex is essential to reach superior screw-bone construct stability and stiffness.
Related Concept Videos
Residual Stresses in Circular Shafts
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Plastic Deformation in Circular Shafts
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Stability of structures
Screw: Problem Solving
To evaluate the...