Background:
In a clinical diagnostic laboratory, we evaluated the applicability of the Ion Proton sequencer for screening 409 cancer-related genes in solid tumours.
Methods:
DNA was extracted from formalin-fixed, paraffin-embedded (FFPE) tissue biopsy specimens of 55 solid tumours (20 with matched normal tissue) and four cell lines and screened for mutations in 409 genes using the Ion Proton system. The mutation profiles of these samples were known based on prior testing using the Ion Torrent Personal Genome Machine (46-gene hotspot panel), Sanger sequencing, or fluorescence in situ hybridisation (FISH). Concordance with retrospective findings and additional mutations were evaluated. Assay sensitivity and reproducibility were established. Gene copy number variations (CNVs) detected were confirmed by molecular inversion probe (MIP) array.
Results:
The average Ion Proton (409-gene panel) sequencing output per run was 8 gigabases with 128 million sequencing reads. Of the 15,992 amplicons in the 409-gene panel, 90% achieved a minimum average sequencing depth of 100X. In 59 samples, the Ion Proton detected 100 of 105 expected single-nucleotide variants (SNVs) and all expected deletions (n=8), insertions (n=5), and CNVs (n=7). Five SNVs were not detected due to failed amplification of targeted regions. In 20 tumours with paired normal tissue, Ion Proton detected 37 additional somatic mutations, several in genes of high prognostic or therapeutic significance, such as MET, ALK, TP53, APC, and PTEN. MIP array analysis confirmed all CNVs detected by Ion Proton.
Conclusions:
The Ion Proton (409-gene panel) system was found to be well suited for use in a clinical molecular diagnostic laboratory. It can simultaneously screen 409 genes for a variety of sequence variants in multiple samples using a low input of FFPE DNA with high reproducibility and sensitivity.
More Related Videos
11:15Next Generation Sequencing for the Detection of Actionable Mutations in Solid and Liquid Tumors
Published on: September 20, 2016
13:24Integration of Wet and Dry Bench Processes Optimizes Targeted Next-generation Sequencing of Low-quality and Low-quantity Tumor Biopsies
Published on: April 11, 2016
Related Concept Videos
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Next-generation Sequencing
Next-Generation Sequencing Methods
Although all next-generation methods use different technologies, they all share a set of standard features....
Cancer
Cancers Originate from Somatic Mutations in a Single Cell
Cancers Originate from Somatic Mutations in a Single Cell