Related Experiment Video
Updated: Apr 22, 2026

Preparation and Pathogen Inactivation of Double Dose Buffy Coat Platelet Products using the INTERCEPT Blood System
Published on: December 7, 2012
Comparative Analysis of Various Aspects of Plateletpheresis on the Fenwal Amicus and Fresenius COM.TEC Cell Separator
Joseph Philip1, Amit Kumar Biswas2, Tathagata Chatterjee1
1Department of Immunohaematology and Blood Transfusion, Armed Forces Medical College, Pune, India.
Objective:
To compare the Fenwal Amicus and the Fresenius COM.TEC apheresis instruments regarding donor peripheral blood parameters, operational variables of the instruments, and quality control parameters of the product obtained.
Methods:
We performed 100 platelet collections from 100 voluntary donors using the 2 studied devices. We measured platelet count using an automated analyzer and analyzed the activation statuses using a flow cytometer.
Results:
The median time needed to perform the procedures was significantly longer with the COM.TEC. However, the product we obtained using the Amicus instrument showed higher degrees of platelet-activation. All products we obtained with both instruments had white blood cell counts of less than 5 × 10(6) per bag. We observed no statistical difference regarding collection efficiency and collection rates between the devices.
Conclusion:
Both instruments collected platelets efficiently, with minimal donor discomfort. Compared with the COM.TEC instrument, the Amicus reached the platelet target yield more quickly; however, it displayed an increase in platelet activation.
More Related Videos
08:34Production of Autologous Platelet-Rich Plasma for Boosting In Vitro Human Fibroblast Expansion
Published on: February 24, 2021
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017